ASTMD 2299&EN149 dual-channel particulate filtration efficiency tester

Short Description:

1.Equipment introduction:

Used for rapid and accurate detection of various flat materials, such as glass fiber, PTFE, PET, PP melt-blown composite of a variety of air particulate filter materials resistance, efficiency performance.

 

Product design meets the standards:

GB 2626-2019 respiratory protection, self-priming filter anti-particulate respirator 5.3 filtration efficiency;

GB/T 32610-2016 Technical Specification for Daily protective masks Appendix A Filtration efficiency test method;

 GB 19083-2010 Technical requirements for medical protective masks 5.4 Filtration efficiency;

YY 0469-2011 Medical surgical Masks 5.6.2 Particle filtration efficiency;

GB 19082-2009 Medical disposable protective clothing technical requirements 5.7 Filtration efficiency;

EN1822-3:2012,

EN 149-2001,

EN14683-2005

EN1822-3:2012 (High Efficiency Air Filter – Flat filter media test)

GB19082-2003 (Medical disposable protective clothing)

GB2626-2019 (Self-priming filter anti-particulate respirator)

YY0469-2011 (Surgical Mask for Medical use)

YY/T 0969-2013 (Disposable Medical Mask)

GB/T32610-2016 (Technical specification for Daily protective Masks)

ASTM D2299—— Latex Ball Aerosol test

 


  • FOB Price: US $0.5 - 9,999 / Piece (Consult a sales clerk)
  • Min.Order Quantity: 1Piece/Pieces
  • Supply Ability: 10000 Piece/Pieces per Month
  • Product Detail

    Product Tags

    2 .Security

    2.1 Safety Specifications

    The equipment shall be operated in accordance with the standard operating codes for electrical use and experiments

    2.2 Electrical

    In case of emergency, you can unplug the power supply and disconnect all power supplies. The instrument will be powered off immediately and the test will stop.

     

    3.Technical parameter:

    1)Pressure: 0.4Mpa gas supply pressure

    2)Flow rate: 32L/min, 85L/min, 95L/min

    3)Humidity: 30% (±10)

    4)Temperature: 25℃ (±5)

    5)Test flow range: 15-100L/min

    6)Test efficiency range: 0-99.999%

    7)Average particle size of sodium chloride aerosol – 0.6 μm;

    8)Sodium chloride aerosol concentration – (8±4) mg/m3;

    9) Average particle size of paraffin oil aerosol – 0.4 μm;

    10)Sodium chloride aerosol concentration – (20±5) mg/m3;

    11)Minimum aerosol particle size – 0.1 μm;

    12)Continuous air flow rate from 15 to 100 dm3/min;

    13)Indication of permeability of anti-aerosol elements in the range from 0 to 99.9999%.

    14) Fully automated process of determining the resistance of the filter material at a set air flow;