Compliant with international testing standards:
ASTM G 153, ASTM G 154, ASTM D 4329, ASTM D 4799, ASTM D 4587, SAEJ2020, ISO 4892 All current UV aging test standards.
Technical parameter:
1.Studio size: mm (D×W×H) 450×1170×500
2.Dimensions: mm (D×W×H) 600×1310×1350
3.Center distance of lamp: 70mm
4.The distance between the specimen and the nearest parallel surface of the lamp surface: about 50mm
5.Wavelength range: UV-A wavelength range is 315~400nm
6.Radiation intensity: 1.5W/m2/340nm
7.Temperature Resolution: 0.1℃
8.Lighting temperature range: 50℃~70℃/temperature tolerance is ±3℃
9.Condensing temperature range: 40℃~60℃/temperature tolerance is ±3℃
10.Black panel thermometer measuring range: 30~80℃/tolerance is ±1℃
11.Temperature control method: PID self-tuning temperature control method
12.Humidity range: about 45%~70%R.H (light state)/98% or more (condensation state)
13.Sink requirements: the water depth is not more than 25mm, and there is an automatic water supply controller
14.Standard specimen size: 75×150mm 48pcs
15.Recommended use environment for the instrument: 5~35℃, 40%~85%R·H, 300mm away from the wall