VST definition: The sample is placed in a liquid medium or a heating box, and the temperature of the standard press needle is determined when it is pressed into 1mm of the sample cut from the pipe or pipe fitting under the action of (50+1) N force under the condition of constant temperature rise.
Definition of thermal deformation (HDT) : The standard sample is subjected to a constant three-point bending load in a flat or side-standing manner, so that it generates one of the bending stresses specified in the relevant part of GB/T 1634, and the temperature is measured when the standard deflection corresponding to the specified bending strain increment is reached under the condition of constant temperature rise.
Model number |
YY-300B |
Sample rack extraction method |
Manual extraction |
Control mode |
7 inch touchscreen moisture meter |
Temperature control range |
RT~300℃ |
Heating rate |
A Speed:5±0.5℃/6min;B speed:12±1.0℃/6min。 |
Temperature accuracy |
±0.5℃ |
Temperature measuring point |
1pcs |
Sample station |
3 working station |
Deformation resolution |
0.001mm |
Deformation measuring range |
0~10mm |
Sample support span |
64mm、100mm(Us standard adjustable size) |
Accuracy of deformation measurement |
0.005mm |
Heating medium |
Methyl silicone oil; Flash point above 300℃, below 200 kris (customer’s own) |
Cooling method |
Natural cooling above 150℃, water cooling or natural cooling below 150℃; |
Instrument size |
700mm×600mm×1400mm |
Required space |
Front to back :1m,left to right:0.6m |
Power source |
4500VA 220VAC 50H |