IV.Main components:
(1)YY710 dry floc generator twisting device (only introduced twisting device, twisting box and air collector);
1. Overall size: main machine: 300×1150×350mm;
2. Cutting template: 285×220mm;
3. Disc diameter: Φ82.8㎜;
4. Reciprocating movement: 60 times/min;
5. Rotation Angle: 180 degrees/time (clockwise, counterclockwise alternately);
6.Starting distance: 188±2㎜ (between two disks);
7. Moving disc stroke: 120±2㎜ (straight line);
8. Online interface;
9. Power supply: 220V10 50Hz5;
10. Maximum power consumption: 500W;
11. Weight: 30kg;
(2) Laser dust particle counter
Large flow dust particle counter is a new type of dust particle counter, which is used to measure the
size and number of dust particles in the unit volume of air in a clean environment. It can complete a cubic meter fast and accurate sampling in 36 minutes, which can quickly verify the cleanliness level of the clean room and print out the relevant test results. Fully comply with ISO14644 and 2010 GMP standards. The product uses a very friendly user interface, high resolution large size color touch screen, easy to operate, simple to use. At the same time, the instrument can simultaneously monitor the number of particles and their changes in eight particle size channels, storing data up to 10×500 groups. It is typically used in particulate matter testing, air particle research, indoor air quality assessment, and also for filter performance testing and cleanliness evaluation.
V.Product features:
1. Flow rate: 28.3 L/min;
2. Display eight diameter channel at the same time;
3. Test results output: ISO14644 and 2010 GMP;
4. Large capacity data storage (10×500 sets of data);
5. 7 inch high resolution color touch screen;
6. Can set each purification level of the standard alarm;
7. Optional RS485 communication interface and software;
8. Optional rechargeable battery, can be continuously measured for more than 4 hours;
9.Low power display (optional rechargeable battery);
10.Printer: (thermal/needle);
VI.Technical parameters:
1.Detection range: 100 ~ 300,000 (GMP A, B, C, D);
2.Eight channels: 0.3, 0.5, 0.7, 1.0, 2.0,3.0, 5.0, 10.0, (μm);
0.3,0.5,0.7,1.0,5.0,10.0,15.0,25.0 (μm)(Options)
3.Sampling period: 1 second –59 minutes 59 seconds optional;
3.1, Self-cleaning time: ≤10min;
3.2. Working environment: temperature 10 ~ 35℃, relative humidity ≤75%;
3.3. Vacuum source: built-in air pump, flow control;
4. Laser particle counter flow rate :28.3L/min,±5%;
5. Light source and life: semiconductor laser > 30000h;
6. The maximum allowable sampling concentration: 35,000 PCS/L;
7. High efficiency filter filtration efficiency: better than 99.99%;
8. Power consumption: 145W;
9. Weight: 9.1Kg;
10. Power supply: 220V ± 10%, 50Hz ± 2Hz/built-in lithium battery (optional);
VII.Configuration List:
Item | Name | Qty |
1 | Particle Counter | 1 set |
2 | Sample cutting template | 1 pcs |
3 | Constant velocity intake probe and adapter | 1 set |
4 | Flexible tube | 1 pcs |
5 | On-line software CD | 1 pcs |
6 | Desktop PC(Lenovo) | 1 set |
7 | Sample mounting clamps | 1 pcs |
8 | Particle counter record paper | 1 roll |
9 | High efficiency air particle filter | 1 Set |