YYJ267 Bacterial Filtration Efficiency Tester

Short Description:

Instrument use:

It is used to detect the bacterial filtration effect of medical masks and mask materials quickly, accurately and stably. The design system based on the working environment of the negative pressure biosafety cabinet is adopted, which is safe and convenient to use and has controllable quality. The method of comparing sampling with two gas channels simultaneously has high detection efficiency and sampling accuracy. The large screen can touch the color industrial resistance screen, and can be easily controlled while wearing gloves. It is very suitable for measurement verification departments, scientific research institutes, mask production and other relevant departments to test the performance of mask bacterial filtration efficiency.

Meeting the standard:

YY0469-2011;

ASTMF2100; 

ASTMF2101;

EN14683;


  • FOB Price: US $0.5 - 9,999 / Piece (Consult a sales clerk)
  • Min.Order Quantity: 1Piece/Pieces
  • Supply Ability: 10000 Piece/Pieces per Month
  • Product Detail

    Product Tags

    Instruments features:

    1.Professional negative pressure biological cabinet safe working environment, to ensure the safety of operators;

    2. High negative pressure working chamber, two-stage high efficiency filter, 100% safe emission;

    3. Adopt two-channel six-level Anderson sampling;

    4.Built-in peristaltic pump, peristaltic pump flow size is adjustable;

    5. Special microbial aerosol generator, bacterial liquid spray flow size can be adjusted, atomization effect is good;

    6. Industrial large color touch screen control, easier operation;

    7. USB interface, support data transfer;

    8. RS232/Modbus standard interface, can achieve external control.

    9. The safety cabinet is equipped with LED lighting, easy observation;

    10. Built-in UV disinfection lamp;

    11. Front switch type sealed glass door, easy to operate and observe;

    12. With SJBF-AS operating software, you can control and data processing through the computer,

    13. Seamless docking laboratory information management system.

     

    Technical Parameters:

    Main parameters Parameter scope Resolution Accuracy
    Sampling flow 28.3 L/min 0.1 L/min ±2%
    Spray flow 8 ~ 10 L/min 0.1 L/min ±5%
    Peristaltic pump flow 0.006~3 mL/min 0.001 mL/min ±2%
    Pressure before sampling flowmeter -20 ~ 0 kPa 0.01 kPa ±2%
    Spray flowmeter front pressure 0 ~ 300 kPa 0.1kPa ±2%
    Negative pressure of aerosol chamber -90 ~ -120 Pa 0.1Pa ±1%
    Working temperature 0~50 ℃
    Cabinet negative pressure > 120Pa
    Data storage capacity Scalable capacity
    High efficiency air filter performance ≥99.995%@0.3μm,≥99.9995%@0.12μm
    Two-channel 6-stage Anderson sampler

    Trapped particle size

    Ⅰ>7μm, 

    Ⅱ4.7~7μm,

    Ⅲ3.3~4.7μm,

    Ⅳ2.1~3.3μm ,

    Ⅴ1.1~2.1μm ,    

     Ⅵ0.6~1.1μm

    Total number of positive quality control sampler particles 2200±500 cfu
    Median diameter of aerosol generator mass Average particle diameter (3.0±0.3 µm), geometric standard deviation ≤1.5
    Six-stage Anderson sampler captures particle size Ⅰ>7 µm; 

    Ⅱ(4.7~7 µm); 

    Ⅲ(3.3~4.7 µm); 

    Ⅳ(2.1~3.3 µm); 

    Ⅴ(1.1~2.1 µm); 

    Ⅵ(0.6~1.1 µm)

    Aerosol chamber specifications L 600 x Ф85 x  D 3mm
    Ventilation flow of negative pressure cabinet >5m3/min
    Main engine size Inner:1000*600*690mm  Outer:1470*790*2100mm
    Work noise < 65db
    Working power supply AC220±10%,50Hz,1KW

     

     