Technical parameters and indicators:
1. Temperature control range: room temperature ~ 300℃
2.Heating rate: 120℃/h [(12±1)℃/6min]
50℃/h [(5±0.5)℃/6min]
3.The maximum temperature error: ±0.5℃
4. Deformation measurement range: 0 ~ 3mm
5. The maximum deformation measurement error: ±0.005mm
6.Deformation measurement display accuracy: ±0.01mm
7. Sample rack (test station) : 6 multi-point temperature measurement
8. The sample support span: 64mm, 100mm
9. Load rod and indenter (needle) weight: 71g
10. Heating medium requirements: methyl silicone oil or other media specified in the standard (flash point greater than 300℃)
11. Cooling method: water cooling below 150 ° C, 150 ° C natural cooling or air cooling (air cooling equipment needs to be prepared)
12. With the upper limit temperature setting, automatic alarm.
13.Display mode: LCD Chinese (English) display
14. Can display the test temperature, can set the upper limit temperature, automatically record the test temperature, the temperature reaches the upper limit automatically stop heating.
15. Deformation measurement method: special high-precision digital display table + automatic alarm.
16. With automatic exhaust oil smoke system, can effectively inhibit the emission of oil smoke, always maintain a good indoor air environment.
17. Power supply voltage: 220V±10% 10A 50Hz
18. Heating power: 3kW