YYT1000 Bacterial Filtration Efficiency Detector(BFE)

Short Description:


Product Detail

Product Tags

Product introduction

YYT1000 Bacterial Filtration Efficiency Detector not only conforms to the requirements of medical surgical masks technology YY0469-2011 appendix B in the test method for filter efficiency (BFE) bacteria first b. 1.1.1 test instrument, but also conforms to the American society for testing material ASTMF2100, ASTMF2101, the requirements of the European EN14683 standards, innovative improvements were made on the basis of this, with double pneumatic contrast sampling method at the same time, improve the precision of sampling,It is suitable for measuring and testing departments, scientific research institutes, mask manufacturers and other relevant departments to test the performance of mask bacterial filtration efficiency.

Execution standard

YY0469

Q/0212 ZRB003-2015

ISO/DIS 22611

ASTMF2101

EN14683

Product features

1.Negative pressure test system to ensure the safety of operators;
2.Built - in peristaltic pump in negative pressure cabinet, A, B two - way, six Andersen;The peristaltic pump flow can be set;
3.The flow rate of night spray can be set and the atomization effect is good.
4.Embedded high speed industrial microcomputer control;10.4 inches industrial high brightness color touch screen;
5.USB interface, support U disk data storage;
6.Cabinet built-in high brightness lights;Built-in leakage protection switch to protect the safety of operators;
7.The inner layer of the cabinet is made of stainless steel, the outer layer is sprayed with plastic and cold-rolled, and the inner and outer layers are insulated and flame retardant.
8.The front switch type glass door is convenient for the experimenter to observe and operate.
9.Detachable bracket, adjustable bracket height;Supporting and moving dual purpose casters.

Specifications

Key Specification

Specification Range

Resolution

Accuracy

A Route Sampling Flow

28.3L/min

0.1L/min

Within ± 2.5%

B Route Sampling Flow

28.3L/min

0.1L/min

Within ± 2.5%

Spray Flow

(8~10)L/min

0.1L/min

Within ± 2.5%

Peristaltic Pump Flow

(0.006~3.0)mL/min

0.001ml/min

Within ± 2.5%

Front Pressure of A Route

Sampling Flowmeter

(-20~0)kPa

0.01kPa

Within ± 2.5%

Front Pressure of B Route

Sampling Flowmete

(-20~0)kPa

0.01kPa

Within ± 2.5%

Front Pressure of

Spray Flowmeter

(0~300)kPa

0.01kPa

Within ± 2.5%

Ambient Temperature

(-40~99)℃

0.1℃

Within ± 2.5%

Negative Pressure of the Aerosol

Chamber

(-90~-120)Pa

0.1Pa

Within ±2.0%

Negative Pressure of the Cabinet

-50~-200pa

 

Data Storage Ability

>100000 sets

 

Vortex Mixer Test Tube

Specification and Quantity

Φ16×150mm test tube,8pcs

High Efficiency Particulate Air

Filter Properties

Filter Efficiency≥99.99% for particles> 0.3μm

Median diameter of aerosol  generator mass

Average diameter: (3.0±0.3)μm; Geometric Standard

generator mass Deviation≤1.5

 

Double Routes 6-lever Andersen Sampling apparatus

Level1>7μm; Level 2: (4.7~7)μm;

Level 3: (3.3~4.7)μm; Level 4: (2.1~3.3)μm;

Level 5: (1.1~2.1)μm; Level 6: (0.6~1.1)μm

Size of Aerosol Chamber

 

600×85×3mm (Length×Diameter×Thickness)

 

Total number of positive quality

control sampler particles

(2200±500)cfu

 

Negative Chamber Flow Speed

≥5m3/min

 

Size of Negative Pressure

Chamber Door(W×D)

1000×730mm

Size of Main Machine(W×D×H)

1180×650×1300mm

 

Size of Supporting

Frame(W×D×H)

1180×650×600mm, Height within 10cm is adjustable

Noise of the Tester

<65dB(A)

Weight

150kg

Power Supply

AC220V±10%, 50Hz

Power Consumption

<1500W


  • Previous:
  • Next:

  • Write your message here and send it to us