(china )YYT139 Total Inward Leakage Tester

Short Description:


Product Detail

Product Tags

INTRODUCTION

The Inward Leakage Tester is used to test the leakage protection performance of respirator and protective clothing against aerosol particles under certain environmental conditions.

The real person wears a mask or respirator and stands in the room (chamber) with a certain concentration of aerosol (in the test chamber). There is a sampling tube near the mouth of the mask to collect the aerosol concentration in the mask. According to the requirements of the test standard, the human body completes a series of actions, reads the concentrations inside and outside the mask respectively, and calculates the leakage rate and overall leakage rate of each action. The European standard test requires the human body to walk at a certain speed on the treadmill to complete a series of actions.

Protective clothing test is similar to the test of mask, requires real people to wear protective clothing and enter the test chamber for a series of tests. The protective clothing also has a sampling tube. The aerosol concentration inside and outside the protective clothing can be sampled, and clean air can be passed into the protective clothing.

Testing Scope:

Particulate Protective Masks, Respirators, Disposable Respirators, Half Mask Respirators, Protective Clothing, etc.

Testing Standards:

GB2626(NIOSH) EN149 EN136 BSEN ISO13982-2

SAFETY

This section describes the safety symbols that will appear in this manual. Please read and understand all precautions and warnings before using your machine.

  HIGH VOLTAGE! Indicates that ignoring the instructions may lead to an electric shock hazard for the operator.
  NOTE! Indicates operational hints and useful information.   
  WARNING! Indicates that ignoring the instructions may damage the instrument.

SPECIFICATION

Test Chamber:
Width 200 cm
Height 210 cm
Depth 110 cm
Weight 150 kg
Main Machine:
Width 100 cm
Height 120 cm
Depth 60 cm
Weight 120 kg
Electric and Air Supply:
Power 230VAC, 50/60Hz, Single Phase
Fuse 16A 250VAC Air Switch
Air Supply 6-8Bar Dry and Clean Air, Min. Air Flow 450L/min
Facility:
Control 10” Touchscreen
Aerosol Nacl, Oil
Environment:

Voltage Fluctuation

±10% of rated voltage

BRIEF INTROCUCTION

dfgh
jklfhg

Treadmill Power Socket1

Power Switch for Test Chamber                 Treadmill Power Socket

Treadmill Power Socket2

Exhaust Blower on Bottom of Test Chamber

Treadmill Power Socket3

Sampling Tubes Connection Adaptors inside Test Chamber

Connection Methods refers to Table I

Be sure D and G with plugs on it when operating the tester.

Treadmill Power Socket4

Samples Tubes for Masks (Respirators)

Treadmill Power Socket5

Treadmill Power Socket6
Treadmill Power Socket7

Sampling Tubes

Treadmill Power Socket8

Plugs for connecting the sampling tube connectors

Touchscreen Introduction

Treadmill Power Socket9

Click the button below to select GB2626 Nacl, GB2626 Oil, EN149, EN136 and other mask test standards, or EN13982-2 protective clothing test standard.

English/中文:Language Selection

GB2626Salt Testing Interface:

Treadmill Power Socket10

GB2626 Oil Testing Interface

Treadmill Power Socket10

EN149 (salt) test interface:

Treadmill Power Socket11

EN136 Salt Testing Interface

Treadmill Power Socket13

Background Concentration:the concentration of particulate matter inside the mask measured by a real person wearing a mask (respirator) and standing outside the test chamber without aerosol;

Environmental Concentration:the aerosol concentration in the test chamber during the test;

Concentration In the Mask:during the test, the aerosol concentration in the mask of the real person after each action;

Air Pressure in the Mask:the air pressure measured in the mask after wearing the mask;

Leakage Rate:the ratio of aerosol concentration inside and outside the mask measured by a real person wearing a mask;

Test Time:Click to start the test timing;

Sampling Time:Sensor Sampling Time;

Start / Stop:start the test and pause the test;

Reset:Reset the test time;

Start Aerosol: after selecting the standard, click to start the aerosol generator, and the machine will enter the preheating state. When the environmental concentration reaches the concentration required by the corresponding standard, the circle behind the environmental concentration will turn green, indicating that the concentration has been stable and can be tested.

Background Measurement: background level measurement;

NO 1-10: the 1st-10th human tester;

Leakage rate 1-5: leakage rate corresponding to 5 actions;

Overall leakage rate: the overall leakage rate corresponding to five action leakage rates;

Previous / next / left / right: used to move the cursor in the table and select a box or the value in the box;

Redo: select a box or the value in the box and click redo to clear the value in the box and redo the action;

Empty: clear all data in the table (Be sure you have written down all the data).

Back: return to the previous page;

EN13982-2 Protective Clothing (salt) Test interface:

Treadmill Power Socket14

A in B out,B in C out,C in A out:Sampling methods for different air inlet and outlet modes of protective clothing;


  • Previous:
  • Next:

  • Write your message here and send it to us